3 coefficient k(2), 4 coefficient k(4) – Metrohm viva 1.0 (process analysis) User Manual
Page 639
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
627
5.6.7.4.3
Coefficient k(2)
Variable name
RS.’Command name’.’Substance name’.CALC2
Description
Second-order coefficient of the calibration curve or standard addition curve.
Decimal places
0…3…6
Unit
Concentration unit = #g/L
#A L
² g⁻² (for Evaluation quantity
= Height)
#C L
² g⁻² (for evaluation quantity
= Area)
Concentration unit = #L/L
#A (for Evaluation quantity =
Height)
#C (for Evaluation quantity =
Area)
Concentration unit = #mol/L
#A L
² mol⁻² (for Evaluation quan-
tity = Height)
#C L
² mol⁻² (for Evaluation quan-
tity = Area)
Remarks
The y axis of the calibration curve is dimensionless because of the fact that it
is standardized in the Dilution Titration (DT) calibration method. The effective
addition volume VEFF is always applied to the x axis. This means that the unit
for the coefficient k(2) = L
⁻².
5.6.7.4.4
Coefficient k(4)
Variable name
RS.’Command name’.’Substance name’.CALC4
Description
Fourth-order coefficient of the calibration curve or standard addition curve.
Decimal places
0…3…6
Unit
Concentration unit = #g/L
#A L
⁴ g⁻⁴ (for Evaluation quantity
= Height)
#C L
⁴ g⁻⁴ (for Evaluation quantity
= Area)
Concentration unit = #L/L
#A (for Evaluation quantity =
Height)
#C (for Evaluation quantity =
Area)
Concentration unit = #mol/L
#A L
⁴ mol⁻⁴ (for Evaluation quan-
tity = Height)
#C L
⁴ mol⁻⁴ (for Evaluation quan-
tity = Area)
Remarks
The y axis of the calibration curve is dimensionless because of the fact that it
is standardized in the Dilution Titration (DT) calibration method. The effective
addition volume VEFF is always applied to the x axis. This means that the unit
for the coefficient k(4) = L
⁻⁴.