5.6 Evaluation subwindow
■■■■■■■■■■■■■■■■■■■■■■
580
■■■■■■■■
viva 1.0 (for Process analysis)
5.6.3.2
Substances - Recognition
Subwindow section: Method
▶ Evaluation ▶ Substances ▶ Recognition
Table
The table on the Recognition tab contains an overview of the parame-
ters that define the assignment of the substances to the peaks/steps that
are found. The table cannot be edited directly and contains the following
columns:
Substance
Name of the substance.
Active
Display as to whether the evaluation for this substance is active or not
active.
Characteristic potential
Verification potential for the assignment of a peak to a substance.
Tolerance
Tolerance of the verification potential for the assignment of a peak to a
substance.
Min. width
Minimum peak width for the assignment of a peak to a substance.
Max. width
Maximum peak width for the assignment of a peak to a substance.
Min. measured quantity
Minimum measured quantity for the assignment of a peak to a substance.
Sweep direction
Shows whether the substance is to be evaluated during forward or reverse
movement of the cyclical sweep.
Modifies the sequence of shown substances by moving the selected line
upward.
Modifies the sequence of shown substances by moving the selected line
downward.