Metrohm viva 1.0 (process analysis) User Manual
Page 40
2.3 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
viva 1.0 (for Process analysis)
Designation
Description
Commands
OVS
Value of the output at the start of the command. Direct controller/2-point con-
troller command status: 1 = active, 0 = inactive
PID/PWM: -100 to 100% (value of the analog output or the pulse frequency)
CONTROL
PK{x}.ABS
Absorbance of the peak x (1 - 9) in mAU
MEAS Spec
PK{x}.TRN
Transmission of the peak x (1 - 9) in %
MEAS Spec
PK{x}.CNT
Intensity of the peak x (1 - 9) in counts
MEAS Spec
PK{x}.CNR
Intensity of the peak x (1 - 9) in the reference spectrum in counts
MEAS Spec
PK{x}.CND
Intensity of the peak x (1 - 9) in the dark spectrum in counts
MEAS Spec
PK{x}.WVL
Wavelength of the peak x (1 - 9) in nm
MEAS Spec
PK{x}.SAT
Detector with peak x (1 - 9) saturated: 1 = saturated;0 = not saturated
MEAS Spec
RPM
Stirring rate in min
-1
CVS
RAN
Current absolute rotation angle of the rack in ° in relation to the axis of the
selected tower (entry when exiting the command)
MOVE
RPO
Current rack position (entry when exiting the command); 0 means 'not defined'
MOVE
SAN
Current absolute swing angle of the robotic arm in ° (entry upon exiting the com-
mand)
MOVE, SWING
SDV
Actual value of the standard deviation when the command is canceled
CALL COND
SPO
Current external position (entry when ending the command); 0 means invalid
position
SWING
STC{x}
Concentration of the standard x (1 - 50) used for the command
CAL LOOP Opt
STY
Type of stop with which the command was stopped: 1 = normal; 0 = manual
or after error
CAL MEAS
Opt, CAL
Spec, CVS,
MEAS Opt,
MEAS Opt
Conc, MEAS
Ref, MEAS
Spec, MEAS
TMF
TMF
Transmission factor (dimensionless)
MEAS TMF
TOL
Tolerance within which a measured peak can be assigned to a calibration wave-
length.
CAL Spec
TOU
Timeout status: 1 = Max. waiting time expired; 0 = Max. waiting time not
expired
RECEIVE,
SCAN, TRANS-
FER
UNI
Mapping unit which is defined for an IO port in the configuration
ANALOG IN,
ANALOG OUT