7 prep, 1 prep - overview, Prep – Metrohm viva 1.0 (process analysis) User Manual
Page 505
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
493
imum speed depends on the cylinder volume of the dosing unit used. The
aspiration and expulsion of the sample should take place at a speed of <
10 mL/min.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
CAUTION
The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.
5.5.2.7.7
PREP
5.5.2.7.7.1
PREP - Overview
Dialog window: Method
▶ PREP ▶ Properties... ▶ PREP - 'Command name'
Command for rinsing the cylinder and tubings of a dosing unit.
Instruments
This command can be executed with the following instruments:
Voltammetry: 894
Dosing Interface: 846
Sample changer: 814, 815, 858, 919
Photometer: 089
Appearance
The command has the following appearance:
Parameters
The parameters for the PREP command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the PREP command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':