8 empty, 1 empty - overview, Empty – Metrohm viva 1.0 (process analysis) User Manual
Page 508

5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
496
■■■■■■■■
viva 1.0 (for Process analysis)
5.5.2.7.8
EMPTY
5.5.2.7.8.1
EMPTY - Overview
Dialog window: Method
▶ EMPTY ▶ Properties... ▶ EMPTY - 'Command name'
Command for emptying the cylinder and tubings of a dosing unit.
Instruments
This command can be executed with the following instruments:
Voltammetry: 894
Dosing Interface: 846
Sample changer: 814, 815, 858, 919
Photometer: 089
Appearance
The command has the following appearance:
Parameters
The parameters for the EMPTY command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the EMPTY com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
CYL
Cylinder volume of the dosing unit used for the com-
mand
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended