5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
478
■■■■■■■■
viva 1.0 (for Process analysis)
CAUTION
The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.
Filling rate
Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the buret used and on the diameter of the hose at the
selected fill port.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
5.5.2.7.3
ADD STD
5.5.2.7.3.1
ADD STD - Overview
Dialog window: Method
▶ ADD STD ▶ Properties... ▶ ADD STD - 'Command name'
Command for addition of standard solutions.
Appearance
The command has the following appearance:
Parameters
The parameters for the ADD STD command are configured in the follow-
ing dialog window:
■
Command variables
The following command variables are generated by the ADD STD com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
ACO
Addition counter (Number) = current number of
completed runs of the command
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started