2 control - properties – Metrohm viva 1.0 (process analysis) User Manual
Page 581
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
569
Identification
Description
OVS
Value of the output at the start of the command.
Direct controller/2-point controller command status:
1 = active, 0 = inactive
PID/PWM: -100 to 100% (value of the analog output
or the pulse frequency)
OVE
Value of the output at the end of the command.
Direct controller/2-point controller: 1 = active, 0 =
inactive
PID/PWM: -100 to 100% (value of the analog output
or the pulse frequency)
5.5.2.11.6.2
CONTROL - Properties
Dialog window: Method
▶ CONTROL ▶ Properties... ▶ CONTROL - 'Command
name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Shows the instrument type.
Selection
IO Controller
Default value
IO Controller
Control parameters
Controller description
Selection of the controller output which is to be controlled.