2 swing - properties – Metrohm viva 1.0 (process analysis) User Manual
Page 518
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
506
■■■■■■■■
viva 1.0 (for Process analysis)
Sample changer: 814, 815, 858, 919
Appearance
The command has the following appearance:
Parameters
The parameters for the SWING command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the SWING command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
SAN
Current absolute swing angle of the robotic arm in °
(entry upon exiting the command)
SPO
Current external position (entry when ending the
command); 0 means invalid position
5.5.2.8.3.2
SWING - Properties
Dialog window: Method
▶ SWING ▶ Properties... ▶ SWING - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.