Step 6: do an autotest in smart remote mode – Fluke Networks CertiFiber Pro Users Manual User Manual
Page 65

Chapter 2: Get Started
Tutorial: Certify Fiber Optic Cabling
43
Step 6: Do an Autotest in Smart Remote Mode
Caution
If you disconnected a test reference cord from the
output of the tester or remote, you must set the
reference again to make sure your measurements
are reliable.
6-1 On the SET REFERENCE screen, when the set reference or
TRC verification procedure is completed, tap NEXT to see
how to connect to the link under test.
6-2 Clean and inspect all the connectors.
6-3 Make the connections to do the test on the fiber link, as shown
on the screen and in Figure 13 above Fiber Link Test, then tap
DONE.
6-4 Tap TEST on the main tester or press on the main or
remote tester.
6-5 Halfway through the test, the tester tells you to switch the
input and output fibers for the bi-directional test.
Caution
Switch the connections at both ends of the patch
panel or fiber under test, not at the tester's and
remote’s ports. If you disconnect a test reference
cord from an output port on the tester or remote,
the reference value will be unreliable.