Characteristic impedance – Fluke Networks CertiFiber Pro Users Manual User Manual
Page 400

Versiv Cabling Certification Product Family
Technical Reference Handbook
378
HDTDX Analyzer Specifications for Cables <100 m (328 ft)
The specifications below are typical for cables less than
100 m (328 ft).
HDTDR Analyzer Specifications for Cables <100 m (328 ft)
The specifications below are for HDTDR tests on cables less than
100 m (328 ft) long.
Characteristic Impedance
The tester gives an estimate of the cable's impedance at
4 m from the start of the link. The accuracy of the measurement is
relative to a 100 terminating resistance.
Parameter
Twisted Pair Cable
Distance accuracy
(1 ft (0.3 m) + 2 % distance)
Distance resolution
1 ft or 0.1 m
Parameter
Twisted Pair Cable
Distance accuracy
(1 ft (0.3 m) + 2 % distance)
Distance resolution
1 ft or 0.1 m
Parameter
Twisted Pair Cable
Range
70 - 180
Accuracy
± (5 + 5 % of 100 – Measured)
Resolution
1