Chapter 16: Maintenance

Clean the DI-1000 Video Probe

367

Clean the DI-1000 Video Probe

To clean the case, use a soft cloth that is moist with a mild
detergent.

Caution

To prevent damage to the case, do not use solvents
or abrasive materials.

To clean the lens, remove the adapter tip, then wipe the lens with
an optical-grade cloth that is moist with an optical-grade cleaning
solution.

Options and Accessories

For a complete list of options and accessories go to the Fluke
Networks website at

www.flukenetworks.com

.

To order options and accessories, contact an authorized Fluke
Networks distributor.

The message “Selected test limit or fiber type is not valid for this test”
shows.

 The test limit requires two wavelengths, but you selected only one. Select

the both wavelengths on the TEST SETUP screen.

 The installed module does not support the selected fiber type or test

limit. For example, the test limit has a singlemode fiber type, but a
multimode module is installed. Select a different fiber type or test limit
or install a different module.

Table 19. Possible Solutions for Unusual Behavior (continued)