Certifiber pro olts module specifications, Certifiber pro power meter specifications – Fluke Networks CertiFiber Pro Users Manual User Manual
Page 401

Chapter 17: Specifications
CertiFiber Pro OLTS Module Specifications
379
CertiFiber Pro OLTS Module Specifications
CertiFiber Pro Power Meter Specifications
Input connector
Interchangeable connector adapter
(LC standard)
Detector type
InGaAs
Wavelengths
850 nm, 1310 nm, 1490 nm, 1550 nm, 1625 nm
Power measurement
range
850 nm: +10 dBm to -65 dBm
All other wavelengths: +10 dBm to -70 dBm
Display resolution
dB, dBm: 0.01
Linear (nW, µW, mW): 0.001 to 9.999, 10.00 to 99.99,
100.0 to 999.9 depending upon value
Display update rate
1 reading per second
Power measurement
uncertainty (accuracy)
5 % 32 pW
1,2
Measurement linearity
< 0.1 dB
3
Re-calibration period
1 year
1. 100 pW at 850 nm
2. Under the following conditions:
Power level 100 µW (-10 dBm), continuous wave (CW) for absolute power at
850 nm and 1310 nm.
Divergent beam, NA = 0.20 for 50/125 µm and NA = 0.14 for 9/125 µm.
Ambient temperature 23° ± 1 °C.
SC/UPC connector with ceramic ferrule.
After a 5 minute warm-up.
Traceable to NIST.
3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23 °C ± 1 °C.
After a 5 minute warm-up.