Chapter 10: The FiberInspector Test
How to Do the FiberInspector Test
283
To do the FiberInspector test
1
Connect the probe to the type A US port on the side of the
tester.
2
Make sure the correct tip is on the probe.
3
Clean the connector that you will inspect.
4
Do the FiberInspector test:
To do the FiberInspector test when it is in the project
a. On the home screen, tap the test setup panel.
b. On the CHANGE TEST screen, tap the FiberInspector test,
then tap USE SELECTED.
c. Tap Test or press .
To do the FiberInspector test when it is not in a project
On the home screen, tap the TOOLS icon, then tap
FiberInspector.
5
Insert the probe into the connector.
6
To adjust the focus, turn the knob on the probe clockwise or
counterclockwise.
Note
The button on the DI-1000 probe has no function
when you use the probe with the Versiv tester.
7
To save the image, tap SAVE. Enter an ID for the test, then tap
SAVE.
Notes
The Auto Save function does not operate with the
FiberInspector test.
Continuous use of the video probe decreases the
battery life. To increase the battery life, connect
the ac adapter when you use the probe for more
than a few minutes.