372
Summary of Performance Specifications
Note
All specifications for tests on twisted pair cabling
apply to 100
cable. Contact Fluke Networks for
information on measurement performance for
cable with a different impedance.
For Category 6
A
/Class E
A
test modes or below, the DSX-5000
CableAnalyzer modules are compliant with Level IIIe
requirements of TIA 1152 and Level IV of IEC 61935-1.
For Class F
A
test modes, the DSX-5000 CableAnalyzer modules are
compliant with Level V requirements as in the draft 4
th
edition of
IEC 61935-1.
Length
Note
Length specifications do not include the
uncertainty of the cable’s NVP value.
Parameter
Twisted Pair Cabling
Without Remote
With Remote
Range
800 m (2600 ft)
150 m (490 ft)
Resolution
0.1 m or 1 ft
0.1 m or 1 ft
Accuracy
(0.3 m + 2 %); 0 m to 150 m
(0.3 m + 4 %); 150 m to 800 m
(0.3 m + 2 %)