Chapter 7: How to Certify Fiber Cabling

Requirements for Reliable Fiber Test Results

165

Make sure you select the correct fiber type and test limit for
the job, and the index of refraction for the fiber is correct. See
Table 8 on page 172.

Make sure the battery is fully charged.

Send the modules to a Fluke Networks service center every 12
months for factory calibration.

About the Reference for Fiber Tests

The reference procedure for fiber cable sets a baseline power
level for loss measurements. If the power level that enters the
fiber from the source changes, the reference and your loss
measurements will be incorrect. The power level can change, for
example, when the temperature at the job site increases or
decreases or when you disconnect then reconnect a test reference
cord at the tester’s output port. So, it is important to set the
reference frequently.

When to Set the Reference

Note

At the job site, turn on the testers and let them sit
for a minimum of 5 minutes before you set the
reference. Let them sit longer if they are above or
below ambient temperature.

The tester requires you to set the reference at these times:

When you change the CertiFiber Pro module in the main or
remote tester.

When you use a different remote tester.

When you change the Reference Method in the test setup.

Set the reference also at these times:

At the start of each day, at the job site, then at regular
intervals during the day. For example, set the reference when
you start tests on a different series of fibers.