beautypg.com

Yaskawa MotionWorks IEC Toolboxes User Manual

Page 587

background image

Toolboxes Function Block ErrorID List

MotionWorks IEC61131-3 Toolboxes: 2013-09-13

574

8715

The socket could not be connected.

8716

There is no network routing path to the specified address.

8717

The socket is already connected to another endpoint.

8718

The socket connection attempt was actively refused by the remote peer.

8719

The socket was not connected to a remote endpoint. Call Y_ConnectSocket prior to
Y_ReadDevice or Y_WriteDevice.

8720

An error occurred trying to get or set the device option.

8721

The communication device could not be read.

8722

The communication device could not be written.

8723

The Buffer argument to WriteDevice and ReadDevice is required.

8724

The device option ID was invalid.

8725

The device option value was not the right size or the data was out of range.

8726

The serial port ID was not a valid serial port.

8727

The serial port could not be opened.

Toolbox ErrorIDs

10020

ProductSize cannot be less than or equal to zero

10021

Maximum allowed consecutive missed registration marks reached

10022

Product or circular buffer overrun / full

10023

Buffer size too small / cannot be zero

10024

DataSize must be greater than zero

10025

Might be crossed or the same non-zero value

10026

Positive Position Limit must be greater than Negative Position Limit

10027

Negative Position Limit must be less than Positive Position Limit.

10028

Positive Velocity Limit must be LREAL#0.0 or greater.

10029

Negative Velocity Limit must be LREAL#0.0 or lower.

10030

Positive Acceleration Limit must be greater than 0.

10031

Negative Acceleration Limit must be less than 0.

10032

Positive Deceleration Limit must be greater than 0.

10033

Negative Deceleration Limit must be less than 0.

10034

Interpolation calculation error.

10035

Gripper Close Error (Timeout)

10036

Gripper Open Error (Timeout)

10037

Offset cannot be in the same direction as the original motion into the limit switch.

10038

CamData.LastSegment must be greater than 0 and less than 400, or whatever value has
been declared as the ARRAY size in the CTB_Types file.

10039

Cam Segment 'Resolution' cannot be zero unless the CurveType is
TB_CurveType#StraightLine..

10040

Curve Type selected in a segment is not valid.

10041

Total pairs required would exceed DataType definition for MS_Array_Type based on number
of segments and resolution settings in CamData.

10042

Master must be always increasing from segment to segment.