How to proceed
7.8.3
Load working sample table
1.
Select program part Workplace.
2. In
subwindow
Run
, click on tab
Determination series
.
3.
Click on menu item
Sample table, Load...
in subwindow
Run
.
The window
Load sample table
opens.
4.
Select desired sample table anc click on [Load].
The sample data of the selected sample table is loaded into the working sample
table.
7.8.4
Edit working sample table
1.
Select program part Workplace.
2. In
subwindow
Run
, click on tab
Determination series
.
3.
If desired, load existing sample table into working sample table.
4.
Click on menu item
Sample table, Properties...
in subwindow
Run
.
The window
Properties - Determination series
opens.
5.
Define properties for the determination series and close the window with
[OK]
.
6.
Select desired line to be changed in subwindow
Run
.
7.
Click on menu item
Edit, Edit line
in subwindow
Run
.
The window
Edit line - Working sample table
opens.
8. Select
Method
out of the desired method group, enter
Sample position
, sample
identifications
ID1...ID8
,
Sample size
and
Sample size unit
and click on
[Apply]
.
The current sample data will be entered in the corresponding line of the working
sample table.
9.
Repeat step 8 for each desired determination.
10. Close
window
Edit line - Working sample table
with
[Close]
.
11.
Click on menu item
Sample table,Save as...
in subwindow
Run
.
The window
Save sample table
opens.
12.
Enter desired name in the field
Name
and click on
[Save]
.
The working sample table is saved under the required name as sample table.