Pump - properties – Metrohm tiamo 1.1 Manual User Manual
Page 569

Commands
541
PUMP - properties
Command name
25 characters, [ Pump # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ 814 USB Sample Processor ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Pumps
Tower
[ 1 ], 2
Selection of the tower at the Sample Processor at which the pumps should
be controlled. Both Tower 1 and Tower 2 are always selectable even if the
Sample Processor has got only one tower.
Pump(s)
[ 1 ], 2, 1+2
Selection of the pump that is to be switched. The pump(s) can either be
built-in or externally connected. With
1+2
both pumps at the selected tower
will be switched at the same time.
Action
Switch on
[ on ], off
Switch on the pump(s).
Switch off
on, [ off ]
Switch off the pump(s).
Duration
on, [ off ]; 0 ... [ 10.0 ] ... 9999.9; [ s ], min
Switch on the pump(s) for a particular period.