Subwindow Devices
681
Tower #
Tower parameters
Max. stroke path
0 ... [ 235 ] mm
Entry of the lowest permitted lift position. A lift height of
0 mm
corresponds
to the "Rest position", i.e. the lift is moved right to the top.
Min. beaker radius
1.0 ... 100.0 mm, [ off ]
Definition of the minimum radius which the beakers used on the rack must
have. If the lift is moved to the work position the beaker radii defined in
Rack table for general Sample positions and Special beaker will be com-
pared with the
Min. beaker radius
. If this minimum beaker radius is undercut
then the sequence will be stopped and an error message will appear. With
off
no check will be made.
Lift rate
3 ... [ 25 ] mm/s
Lift speed for manual operation.
Axial distance
100.0 ... 300.0 mm
[ 166.0 mm ]
(778 Sample Processor, 814 USB Sample Processor XL)
[ 196.0 mm ]
(789 Robotic Sample Processor, 815 Robotic USB Sample
Processor XL, 855 Robotic Titrosampler)
Distance between the axis of rotation of the sample rack and turning axis of
the robotic arm.
Swing head
This parameter is only shown when a swing head is mounted on the tower.
Serial number
Shows the serial number of the swing head.
Swing position
[ 0 ] ... 235 mm
Lift position valid for all 4 external positions, at which the swing head turns
to the external positions.
Rinse position
[ 0 ] ... 235 mm