Subwindow Devices
673
Measuring input 1/2
ADC type
Shows the type of analog-digital converter.
ADC serial number
Shows the serial number of the measuring input.
Temperature sensor
[ Pt 1000 ], NTC
Selects the type of temperature sensor connected to the measuring input.
For an
NTC
type temperature sensor two further parameters are visible:
R (25 °C)
1000 ... [ 30000 ] ... 99999 Ohm
Nominal resistance of connected NTC sensor.
B value
1000 ... [ 4100 ] ... 9999
Material constant of the NTC resistance referred to measuring the resis-
tance at 25 °C and 50 °C.
Note
Titrando models 2.8xx.0010 have only one measuring input.
Tower #
Tower parameters
Max. stroke path
0 ... [ 235 ] mm
Entry of the lowest permitted lift position. A lift height of
0 mm
corresponds
to the "Rest position", i.e. the lift is moved right to the top.
Min. beaker radius
1.0 ... 100.0 mm, [ off ]
Definition of the minimum radius which the beakers used on the rack must
have. If the lift is moved to the work position the beaker radii defined in
Rack table for general Sample positions and Special beaker will be com-
pared with the
Min. beaker radius
. If this minimum beaker radius is undercut
then the sequence will be stopped and an error message will appear. With
off
no check will be made.