Commands

439

Sensor

Measuring input

[ 1 ] ... 2

(835, 836, 842)

[ 1 ]

(855)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, pH electrode, [ Metal electrode ], ISE electrode, Conductivity sen-
sor

Selection of the sensor from the list of sensors defined in the sensor table.

For pH and ISE electrodes, the calibration data will be adopted for the de-
termination.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured

continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Titration pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only Titrando)

If this option is enabled, the stirrer will be switched off automatically when

the command is finished.

Switch on/off automatically

[ on ], off

only 751, 799)

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of

the command.

Start conditions

Screenshot and parameters: see STAT pH - Start conditions