4 options, General – Metrohm tiamo 1.1 Manual User Manual
Page 654

626
Configuration
6.3.4 Options
Under
Tools, Options
general program properties can be set on the following 3
tabs:
•
General
Selection of the dialog language and switches the emergency stop
button on/off.
•
Save
Saves settings on exiting tiamo.
•
Settings for pdf files
General
Dialog language
Dialog language
Deutsch, [ English ]
Selection of the dialog language.
Note
For the altered setting to become effective tiamo must be restarted.
Emergency stop button
on, [ off ]
If this option is enabled then the emergency stop button will be shown in all
program parts.
This button can be moved to any position with the left-hand mouse key
pressed down and is shown right at the top on the screen both in the tiamo
program window and outside it on the Windows desktop. A mouse click on
this button immediately stops all running determinations on all active work-
places. The emergency stop is effective even if no user is logged in, e.g.
when the user has been logged out automatically (see Login/Password pro-
tection).
Note
The emergency stop stops all commands and tracks on all workplaces but no
device functions switched on by a command (e.g. pumps, valves, stirrers). To
stop these devices a separate track containing the commands to stop the de-
sired functions must be called in the Exit track with the condition
SV.STO = 1
. The
emergency stop has no effect on actions still running in the manual control.
These actions must be stopped there one by one with the
button.