Commands

417

Temperature

-170.0 ... [ 25.0 ] ... 500.0 °C

Manually entered titration temperature.

Time interval measuring point

[ 1 ] ... 999999 s

Time interval for entering a measuring point in the list of measuring points.

Stop conditions

Condition for stopping the titration.

Command name

25 characters, [ KFC # ]

Name of the command.

Stop time

0 ... 999999 s, [ off ]

Stops when the preset time has elapsed since the start of the titration.

off

means no stop.

Conditioning

Command name

25 characters, [ KFC # ]

Name of the command.

Conditioning

[ on ], off

Switching conditioning on/off. If conditioning is deactivated all following pa-
rameters are not visible.

Start drift

1 ... [ 20 ] ... 999 µg/min

The determination can only be started when the measured drift is below the
entered value. The Status in the

Run

window (Single determination or De-

termination series) then switches to

COND READY

.