beautypg.com

Modify remark, Modify autostart counter – Metrohm tiamo 1.1 Manual User Manual

Page 155

background image

Subwindow Run

127

method can be switched on and off manually here (only in the status

READY

).

The first field after

Statistics

shows the number of determinations that have

already been carried out for the statistics (actual counter). This field cannot
be edited, but in the status

READY

the context-sensitive menu item

Delete

statistics

can be used to reset it to

0

, which at the same time deletes the

statistics data that has already been calculated.
The second field after

Statistics

shows the number of determinations to be

carried out for the statistics (set counter). It contains the value for the

Num-

ber of single determinations

defined in the

START

command as standard.

This field can only be edited in the status

READY

.

Modify remark

With the context-sensitive menu item

Modify remarks

the dialog window

Live modifi-

cations

opens, in which the remark can be edited live during a running determina-

tion.

Remark

100 characters

Freely definable comments about the determination that are saved together

with the determination. Instead of a manual entry it is also possible to select
from prepared Text templates.

Modify autostart counter

With the context-sensitive menu item

Modify autostart counter

the dialog

window

Live modifications

opens, in which the autostart counter can be al-

tered live during a running determination.

Autostart

1 ... 999, [ Sample table ]

Number of automatic, internal starts of determinations within a sample se-
ries.
The first non-editable field after

Autostart

shows the number of determina-

tions that have already been carried out (actual counter).
The second editable field after

Autostart

shows the total number of determi-

nations to be carried out per series (set counter).