Potentiometric evaluation – Metrohm tiamo 1.1 Manual User Manual
Page 375

Commands
347
Stop time
0 ... 999999 s, [ off ]
(only Titrando)
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Filling rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando)
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed with which the buret is to be refilled after the titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
unit used.
Potentiometric evaluation
Parameters for the potentiometric evaluation of the titration curve with automatic
equivalence point recognition. One of the following evaluation methods can be
chosen:
Command name
25 characters, [ DET U # ]
Name of the command.
Evaluation without window
Parameters: see DET pH - Potentiometric evaluation
Evaluation with measured value window (U)
on, [ off ]
With this option set up to nine regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within these
windows and additionally meet the parameters defined for each window will
be recognized. Per window only one EP will be recognized. The defined
windows with their parameters are shown in the table and can be edited
with the following buttons:
Opens the dialog window
Measured value window #
(see below) in which the
parameters for a new window can be entered.
Opens the dialog window
Measured value window #
(see below) in which the
parameters for a new window can be edited.