Commands

441

Min. rate

0.01 ... [ 5.00 (50 µL/min)] ... [ 10.00 (100 µL/min) ] ... [ 40.00 (500 µL/min) ]
... 9999.00 µL/min

Speed of addition at the start and finish of the titration. The slower the

minimum rate that is selected, the slower the titration and the higher the ac-
curacy. the longer it takes until the control point is reached for the first time.
The following rule of thumb should be considered in order to obtain as con-
stant a dosing at the control point as possible:

Min. rate

in µL/min = Ex-

pected rate of reaction in µL/min / 10.

Note
Outside the control range the dosing rate is mainly characterized by the

Max.

rate

. The parameters

Max. rate

and

Dynamics

should be optimized together in

such a way that the titration does not overshoot too much when reaching the
control point. The control range should be set so that the measured value re-
mains within this range while it is kept constant. A slightly larger control range
should be defined for slow reactions (e. g. U = 180 mV). A SET pretitration is

frequently used to reach the control point before the substrate is added. This
means that the STAT titration can be started with a small variation from the con-
trol point. Within the control range the dosing rate is mainly characterized by the

Min. rate

.

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ STAT U # ]

Name of the command.

Titration direction

Selection of the titration direction.

+

Positive measured value alteration.

[ - ]

Negative measured value alteration.

auto

The titration direction is determined automatically from the start measured
value and the set control point.