Metrohm tiamo 1.1 Manual User Manual
Page 564

536
Method
Target
Tower
[ 1 ], 2
Selecting the tower of the Sample Processor for moving to the required po-
sition. Both Tower 1 and Tower 2 are always selectable even if the Sample
Processor has got only one tower.
Move
Selecting the target position on the rack:
[ Sample position ]
Position defined in the Run window as parameter
Sample position
.
Rack position
Position on the rack.
Special beaker
Special position on the rack defined in the rack properties (see Rack prop-
erties - Special beaker).
Relative angle
(only for 778, 789, 814, 815)
The sample rack can be rotated by a certain amount relatively to the current
position independently from rack positions. This can be used for example
to remove vial caps automatically.
Number
[ 1 ] ... 999
(for
Move
=
Rack position
)
[ 1 ] ... 16
(for
Move
=
Special beaker
, only for 778, 789, 814, 815, 855)
[ 1 ] ... 8
(for
Move
=
Special beaker
, only for 730, 774)Specification of the
number of the rack position. This parameter is only visible for
Move
=
Rack
position
or
Special beaker
. The rack position can be entered as a formula as
well. This is necessary for example for the automatic calibration of elec-
trodes. The Formula editor (see Formula editor) is opened by right clicking
into the input field.
Move angle
-259.90 ... [ 5.00 ] ... 259.90 °
This parameter specifies the angle about which the sample rack should be
rotated relatively to the current position. It is only visible for
Move
=
Relative
angle
.
Beaker test
Selection of the action to be carried out if the
Beaker sensor
defined for the
rack doesn't detect a sample at the target position. If the
Robotic arm
is de-
fined as beaker sensor, a Swing Head must be installed and a suitable
Work position
with beaker contact must be defined for the lift. This position
will be moved to automatically for beaker recognition after the command
MOVE
. This section is not visible for
Move
=
Relative angle
.
Display message
[ on ], off
A message is displayed and the determination is stopped.
Stop determination
on, [ off ]
The determination is stopped and the next determination of the series is
started.
Stop determination and series
on, [ off ]
The determination as well as the series are stopped. An Error track will be
started if there is one defined.
Parameters
Shift rate
3 ... [ 20 ] °/s
Entry of the speed at which the sample rack rotates.