Potentiometric evaluation
Parameters for the potentiometric evaluation of the titration curve with automatic
equivalence point recognition. One of the following evaluation methods can be
chosen:
Command name
25 characters, [ DET pH # ]
Name of the command.
Evaluation without window
[ on ], off
With this option set the two following parameters will be applied on the
whole range of the titration curve.
EP criterion
0 ... [ 5 ] ... 200
Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.
EP recognition
Filter for the recognition of equivalence points:
[ all ]
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steepest
jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
(only Titrando)
Only equivalence points with a positive slope of the titration curve will be
recognized.
descending
(only Titrando)
Only equivalence points with a negative slope of the titration curve will be
recognized.
off
The equivalence point evaluation is switched off.