Potentiometric evaluation – Metrohm tiamo 1.1 Manual User Manual
Page 393

Commands
365
Potentiometric evaluation
Parameters for the potentiometric evaluation of the titration curve with automatic
equivalence point recognition. One of the following evaluation methods can be
chosen:
Command name
25 characters, [ DET Upol # ]
Name of the command.
Evaluation without window
Parameters: see DET pH - Potentiometric evaluation
Evaluation with measured value window (Upol)
on, [ off ]
With this option set up to nine regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within these
windows and additionally meet the parameters defined for each window will
be recognized. Per window only one EP will be recognized. The defined
windows with their parameters are shown in the table and can be edited
with the following buttons:
Opens the dialog window
Measured value window #
(see below) in which the
parameters for a new window can be entered.
Opens the dialog window
Measured value window #
(see below) in which the
parameters for a new window can be edited.
Deletes the selected window in the table.
EP criterion
0 ... [ 5 ] ... 200
(only Titrino)
Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion
will not be recognized.