Rack – Metrohm tiamo 1.1 Manual User Manual
Page 717

Subwindow Devices
689
Tower 1 / Tower 2
Max. stroke path
Shows the lowest permitted lift position for the two towers. A lift height of
0
mm
corresponds to the "Rest position", i.e. the lift is moved right to the top.
Beaker sensor
Shows whether the beaker sensor for the two towers is switched on or off.
Each time that a sample position is moved to the beaker sensor checks
whether a beaker is present or not.
Tower 1
Number of pumps
Shows how many pumps are connected to Tower 1.
759 Swing Head
Shows whether a swing head is mounted on Tower 1.
Tower 2
Number of pumps
Shows how many pumps are connected to Tower 1.
Note
A swing head can only be mounted on Tower 1.
Rack
On the tab
Rack
the rack-specific data of the attached rack are shown.
Rack name
Shows the name of the attached rack. If no rack is in position then "
-----
" will
be shown.
Rack code
Shows the rack code of the attached rack. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the
Sample Processor in order to recognize the rack. If no rack is in position
then the display will be empty.
Number of positions
Shows the number of positions on the rack. If no rack is in position then the
display will be empty.
Shift rate
3 ... [ 20 ] °/s
Sample rack speed of rotation for manual operation.