Dosing parameters
The volume to be added and the dosing and filling rate are defined on this tab.
Command name
25 characters, [ Dosing # ]
Name of the command.
Volume
0
.00000 ... [ 10.0000 ] ... 99999.9 mL (Titrando, 846, 814, 815, 855)
0.0001 ... [ 10.0000 ] ... 9999.0 mL
(Titrino)
0.001 ... [ 10.000 ] ... 999.999 mL
(730, 778, 789)
0.01 ... [ 10.000 ] ... 999.999 mL
(774)
Volume to be added.
Dosing rate
[ maximum ]
0.01 ... 166.00 mL/min
(Titrando, 814 815)
0.01 ... 160.00 mL/min
(730, 774, 778, 789)
0.01 ... 150.00 mL/min
(Titrino)
Speed at which dosing is carried out. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing unit used. In case the
entered dosing rate is too high for the selected dosing device, the rate will
be reduced automatically to the highest possible value.
Note
Lower dosing rates should be entered for solutions with higher viscosities.
Filling rate
[ maximum ]
0.01 ... 166.00 mL/min
(Titrando, 846, 814, 815, 855)
0.01 ... 160.00 mL/min
(730, 774, 778, 789)
0.01 ... 150.00 mL/min
(Titrino)
Speed at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing unit used. In case the entered filling rate is too high for the selected
dosing device, the rate will be reduced automatically to the highest possi-
ble value.
Note
Lower filling rates should be entered for solutions with higher viscosities.
Fill automatically at start
[ on ], off
(only Titrando, 846, 814, 815, 855)
The buret is automatically filled before dosing if this option is enabled.
Fill automatically at end
[ on ], off
The buret is automatically filled after dosing if this option is enabled.