Towers – Metrohm tiamo 1.1 Manual User Manual
Page 716

688
Configuration
Data storage under legal control
on, [ off ]
This option must be enabled if the balance has its own data memory (only
for Sartorius balances).
Remarks
1000 characters, [ empty ]
Remarks about the device.
This button is only shown with a balance or barcode reader. In order to de-
lete the device from the device list the device must be disconnected.
Load new program version
If when a device is connected up it is discovered that this has an old program ver-
sion that is not supported by tiamo then it must be updated. This is done by se-
lecting the device in the Device table and using
Edit, Properties
to open the Proper-
ties window for the device. On the tab
General
the button
[Update]
is then active.
Opens the dialog window
Load new program version
.
Old version
Shows the old device program version.
New version
Shows the new device program version to be loaded.
Loads new program version.
Note
Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.
Towers
On the tab
Towers
the tower parameters for Tower 1 and Tower 2 (if present) are
shown for the 730 Sample Changer. These cannot be edited, but must be set on
the device itself via the keyboard.