Stop conditions – Metrohm tiamo 1.1 Manual User Manual

Page 470

background image

442

Method

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.

Measuring point recording

Time interval measuring point

0.1 ... [ 2.0 ] ... 999999 s

(Titrando)

1 ... [ 2 ] ... 999999 s

(Titrino)

Time interval for entering a measuring point in the list of measuring points.

Start time

[ 0 ] ... 999999 s

The measured values are not entered in the list of measuring points until

this time has elapsed since the start of the titration.

Start measured value

-2000.0 ... 2000.0 mV, [ off ]

(Titrando),

-2000 ... 2000 mV, [ off ]

(Titrino)

The measured values are not entered in the list of measuring points until

this value has been reached.

Start rate

0.01 ... 166.00 mL/min, [ off ]

(Titrando)

0.01 ... 150.00 mL/min, [ off ]

(Titrino)

The measured values are not entered in the list of measuring points until
the dosing rate has fallen below this value. This parameter is checked 10 s

after the start at the earliest.

Stop conditions

Screenshot and parameters: see STAT pH - Stop conditions