Stop conditions – Metrohm tiamo 1.1 Manual User Manual
Page 470

442
Method
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement
in the tab
General/Hardware
under
Sensor
is set to
automatic
or
continuous
then the temperature will be meas-
ured continuously.
Measuring point recording
Time interval measuring point
0.1 ... [ 2.0 ] ... 999999 s
(Titrando)
1 ... [ 2 ] ... 999999 s
(Titrino)
Time interval for entering a measuring point in the list of measuring points.
Start time
[ 0 ] ... 999999 s
The measured values are not entered in the list of measuring points until
this time has elapsed since the start of the titration.
Start measured value
-2000.0 ... 2000.0 mV, [ off ]
(Titrando),
-2000 ... 2000 mV, [ off ]
(Titrino)
The measured values are not entered in the list of measuring points until
this value has been reached.
Start rate
0.01 ... 166.00 mL/min, [ off ]
(Titrando)
0.01 ... 150.00 mL/min, [ off ]
(Titrino)
The measured values are not entered in the list of measuring points until
the dosing rate has fallen below this value. This parameter is checked 10 s
after the start at the earliest.
Stop conditions
Screenshot and parameters: see STAT pH - Stop conditions