Metrohm tiamo 1.1 Manual User Manual
Page 730

702
Configuration
Timeout
1000 ... [ 2000 ] ... 9000 ms
Maximum waiting time for receiving characters. If this time is exceeded the
reception will be stopped.
Terminator for send
\0D, \0D\0A, \0D\0A\0A, \0D\0D\0A, [ 'empty' ]
Final character for transmission in hexadecimal code.
Terminator for receive
\0D, \0D\0A, \0D\0A\0A, \0D\0D\0A, [ 'leer' ]
Final character for reception in hexadecimal code.
Code page
[ Cp437 ], Cp850, Cp851
Code page used for the data transmission.
Tests the connection to the balance. The dialog window
Check connection
appears.
Test connection
In order to test the connection to a balance the balance must be selected in the
Device table and then under
Edit, Properties
the Properties window for this device
opened. On the tab
RS 232
the button
[Check connection]
must be pressed.
Opens the dialog window
Check connection
.
If the print key on the connected balance is pressed then the data provided
by the balance will be inserted in the following fields:
Device name
Name of balance.
Sample size
Sample weight.
Sample size unit
Sample weight unit.
ID1
Sample identification ID1.
ID2
Sample identification ID2.
Entry date
Date and time of data entry by balance (only displayed for Sartorius bal-
ances with
Data memory
switched on).
Speichernummer
Internal memory number of the balance (only displayed for Sartorius bal-
ances with
Data memory
switched on).