Metrohm tiamo 1.1 Manual User Manual
Page 451

Commands
423
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(835, 836, 842)
[ 1 ] ... 3
(855)
[ internal D0 ]
(Titrino)
external D1, external D2
(only 736, 751, 799)
Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing.
Solution
24 characters, solution name, [ not defined ]
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If
not defined
is selected no tests will be carried
out.
Tandem dosing
on, [ off ]
(only Titrando)
If this option is enabled, the dosing is carried out without interruption with a
combination of two dosing devices so that the second dosing device is
dosing while the first one is being filled and vice versa (see below).
Dosing device
1 ... [ 2 ] ... 4
(835, 836, 842)
1 ... [ 2 ] ... 3
(855)
Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.
Solution
24 characters, solution name, [ not defined ]
(only Titrando)
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. If
not defined
is selected no tests will be carried out.
Filling rate
0.01 ... 166.00 mL/min, [ maximum ]
(only Titrando)
Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the ex-
change unit or dosing unit used. If the entered filling rate is too high for the
selected dosing device then during dosing it will automatically be reduced
to the largest possible value (see below).
Tandem dosing
In order to ensure continuous dosing, the following points must be taken into ac-
count:
• Keep filling times short
Use the highest possible filling rate in order to keep the filling times as
short as possible. Please note the viscosity and density of the liquid.