Measuring parameters, Evaluations – Metrohm tiamo 1.1 Manual User Manual
Page 508

480
Method
Device type
774 Oven Sample Processor
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Measuring parameters
The measuring parameters are defined on this tab.
Command name
25 characters, [ MEAS T/Flow # ]
Name of the command.
Measuring time
0 ... [ 600 ] ... 999'999 s
Maximum period of time for measurement.
Measuring interval
1 ... [ 10 ] ... 999'999.0 s
Time interval for entering a measuring point in the list of measuring points.
Stop measured value Temp.
50.0 ... 250.0 °C, [ off ]
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Evaluations
The following methods for evaluating measuring curves are defined on this tab.
Command name
25 characters, [ MEAS T/Flow # ]
Name of the command.