Additional measured values, Stat u – Metrohm tiamo 1.1 Manual User Manual
Page 464

436
Method
Threshold value
0.1 ... [ 1.0 ] ... 20.0 pH/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.
Maximum evaluation
on, [ off ]
For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 pH/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.
Additional measured values
Abbildung und Parameter: siehe DET pH - Weitere Messwerte
STAT U
Command for STAT titrations with measured value U.
Devices
This command can be executed with the following devices which have the
STAT
pH
mode:
Titrino: 718, 736, 751, 799
Titrando: 835, 836, 842, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
STAT U
are configured in the following 8 tabs:
•
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
•
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
•
Control parameters
Parameters for the control point.
•
Titration parameters
Parameters for the run of the titration.
•
Stop conditions
Definitions of conditions which cause the stop of the titration.