6 archive audit trail – Metrohm tiamo 1.1 Manual User Manual
Page 794

766
How to proceed
7.2.6
Archive Audit Trail
1.
Select program part Configuration.
2.
Click on symbol
or menu item
Tools, Audit Trail
.
The window
Audit Trail
opens.
3.
Click on menu item
File, Archive
.
The window
Archive Audit Trail
opens.
4.
Enter or select directory and file name of the archive file in the field
Save file as
.
5.
Select the desired option (
All records
or
Records until
including date selection) in
the field
Selection
.
6. Click
on
[OK]
.
The entries selected in the Audit Trail table will be archived as a text file.
Note
Archiving Audit Trail entries is identical to Export, i.e. the Audit Trail entries are
stored in text format. They can no longer be imported back into the Audit Trail
table. The difference to exporting is that the archived entries can be marked in
the column
Archived
and then deleted.
The text files generated by archiving are no longer protected and can be ma-
nipulated. If you want to ensure that these files are archived in an unaltered con-
dition you must use a suitable external backup or archiving program.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)