Additional measured values, Meas t – Metrohm tiamo 1.1 Manual User Manual
Page 501

Commands
473
Slope
0.0 ... [ 0.9 ] ... 10.0
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.
Smoothing factor
2 ... [ 5 ] ... 20
The higher the smoothing factor, the fewer break points will be found.
Window
Measured value, Time, [ off ]
A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.
Lower limit
[ -200.0 ] ... 200.0 µA
(for
Window
=
Measured value
)
[ 0 ] ... 999'999 s
(for
Window
=
Time
)
Measured value for the lower limit of the window.
Upper limit
-200.0 ... [ 200.0 ] µA
(for
Window
=
Measured value
)
0 ... [ 999'999] s
(for
Window
=
Time
)
Measured value for the upper limit of the window.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MEAS T
Command for temperature measurements.
Devices
This command can be executed with the following devices:
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MEAS T
are configured on the following 4 tabs:
•
General/Hardware
Parameters for device, sensor and stirrer.
•
Measuring parameters
Parameters for the measurement.
•
Evaluations
Defining further methods for evaluating measurement curves.
•
Additional measured values
Defining further measured values which can be saved as additional
columns in the list of measuring points.