Commands

473

Slope

0.0 ... [ 0.9 ] ... 10.0

Minimum difference between the slope before and after the break point.

The smaller the difference, the more break points will be found.

Smoothing factor

2 ... [ 5 ] ... 20

The higher the smoothing factor, the fewer break points will be found.

Window

Measured value, Time, [ off ]

A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-

dow. Only the first break point in the defined window will be recognized.

Lower limit

[ -200.0 ] ... 200.0 µA

(for

Window

=

Measured value

)

[ 0 ] ... 999'999 s

(for

Window

=

Time

)

Measured value for the lower limit of the window.

Upper limit

-200.0 ... [ 200.0 ] µA

(for

Window

=

Measured value

)

0 ... [ 999'999] s

(for

Window

=

Time

)

Measured value for the upper limit of the window.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values

MEAS T

Command for temperature measurements.

Devices

This command can be executed with the following devices:

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799

Titrando: 808, 809, 835, 836, 841

Appearance

The command has the following appearance:

Parameters

The parameters for the command

MEAS T

are configured on the following 4 tabs:

General/Hardware

Parameters for device, sensor and stirrer.

Measuring parameters

Parameters for the measurement.

Evaluations

Defining further methods for evaluating measurement curves.

Additional measured values

Defining further measured values which can be saved as additional
columns in the list of measuring points.