Method

Control parameters

The Control parameters can be set for each endpoint separately.

Command name

25 characters, [ SET pH # ]

Name of the command.

Endpoint 1

EP1 at pH

-20.000 ... 20.000, [ off ]

(Titrando),

-20.00 ... 20.00, [ off ]

(Titrino)

Measured value for the first endpoint.

Titration rate

Titration rate

Three predefined sets of parameters can be selected for the titration rate:

slow

,

optimal

and

fast

. For these sets the parameters for the

Control

are not

displayed. In order to edit these parameters the value

user

must be se-

lected.

slow

For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.

[ optimal ]

Parameter set for all standard titrations; optimized for the most frequent

applications.

fast

For less critical rapid titrations.