beautypg.com

Additional measured values, Meas conc – Metrohm tiamo 1.1 Manual User Manual

Page 509

background image

Commands

481

Minimum evaluation Temperature

on, [ off ]

For the minimum measured value the associated time and temperature are

interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 °C/s

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the threshold value.

Maximum evaluation Temperature

on, [ off ]

For the maximum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 °C/s

The evaluation of the maximum begins as soon as the slope of the curve

exceeds the threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values


MEAS Conc

Command for concentration measurements (direct measurement).

Devices

This command can be executed with the following devices:

Titrando: 808, 809, 835, 836

Appearance

The command has the following appearance:

Parameters

The parameters for the command

MEAS Conc

are configured on the following 3

tabs:

General/Hardware

Parameters for device, sensor and stirrer.

Measuring parameters

Parameters for the measurement.

Additional measured values

Defining further measured values which can be saved as additional

columns in the list of measuring points.