490
Method
continuous
A temperature sensor must be connected. The temperature is measured
continuously.
[ automatic ]
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered
Temperature
(see
Meas-
uring parameters
) will be used.
off
The temperature will not be measured. The
Temperature
(see
Measuring pa-
rameters
) which has been entered manually will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
The stirrer is automatically switched off at the end of the command if this
option is enabled.
Standard addition
The standard addition parameters are defined on this tab.
Command name
25 characters, [ STDADD man # ]
Name of the command.
Number of additions
1 ... [ 3 ] ... 19
Number of times the standard solution is to be added.
Volume auxiliary solution
[ 0.000 ] ... 9999.999 mL
Volume of the auxiliary solution (e.g. buffer) to be added to the sample be-
fore the first standard addition.