Titration parameters – Metrohm tiamo 1.1 Manual User Manual
Page 437

Commands
409
Note
You can use the
Delay time
to calculate the maximum
Stop drift
to be used in the
following way: the size of the last added increment depends on the volume of
the attached exchange unit. With a 20 mL exchange unit (10'000 pulses per cyl-
inder volume; for KFT command is independent of the dosing drive) the smallest
possible increment is 2 µL. With a Delay time of 5 s the last 2 µL of reagent must
therefore suffice for 5 s or longer. This gives a drift of ≤ 2 µL/5 s ≤ 24 µL/min.
(The drift may be smaller than 24 µL/min as we do not know whether the last in-
crement would also have been sufficient for 10 s.) This means that if you use a
20 mL exchange unit (10'000 pulses per cylinder volume; independent of the
dosing drive) and 5 s switch-off delay then you should set a value of ≤ 24
µL/min as the stop drift.
Stop time
1 ... 999999 s
(only 720, 736, 751, 758, 784, 795, 799)
The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for
Stop criterion
=
time
and
Delay time
=
inf.
.
Titration parameters
Parameters defining the run of the titration.
Command name
25 characters, [ KFT Ipol # ]
Name of the command.
Titration direction
Selection of the titration direction. If two endpoints are set the direction is al-
ready given and the following selection is ignored.
+
positive measured value alteration
[ - ]
negative measured value alteration
auto
The titration direction is determined automatically from the start measured
value and the set endpoint.
Extraction time
[ 0 ] ... 999999 s
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement
in the tab
General/Hardware
under
Sensor
is set to
automatic
or
continuous
then the temperature will be meas-
ured continuously.