Balance, General – Metrohm tiamo 1.1 Manual User Manual

Page 728

background image

700

Configuration

Action

If the time interval for the GLP test has expired then one of the following actions

will be triggered at both the Start test and the command sequence:

Record message

The message that the GLP test interval has expired will be saved automati-
cally in the determination.

Display message

A message appears and you can then choose whether to continue with the
run or to terminate it. If the run is continued then the message that the GLP

test interval has expired will be saved automatically in the determination.

Stop determination

The running determination will be automatically terminated. The following
message must be confirmed with

[OK]

.

Balance

The parameters for a Balance are set on the following tabs:

General

General device information such as device name, device type, serial
number, etc.

RS232

Selects the serial interface to which the balance is connected.

GLP

Information about GLP test and GLP monitoring.

General

Device name

50 characters, [ device type number_# ]

Name of device (can be freely defined by user).

Device type

Shows the type of device.

Program version

Shows the program version of the device.

Opens the dialog window

Load new program version

. This button is only ac-

tive when the device has an old program version that is not supported by
tiamo which can be updated by tiamo itself.