Commands
565
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Remote box
[ 1 ] ... 4
(only Titrando, 814, 815, 855)
Selection of the remote box at which the lines are to be scanned.
Scan lines
Input signal
Binary pattern of exactly 8 characters (0, 1,*), [ ******** ], signal template
Input of the binary pattern for the expected input signal or selecting a pre-
defined Signal pattern.
The following characters can be entered:
0
= Line inactive
1
= Line active
*
= Any line condition
The bits are always numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0
Example:
*******1
is waiting for an active input on line 0 (1 = set). For
example, this line is set by a Titrino after a titration has been completed and
the Titrino can accept a start signal again.
Note
Input lines that are of no interest or for which no defined condition can be pre-
dicted should also be masked with an asterisk (
*
).
Timeout
on, [ off ]
Maximum waiting time for the input signal.