Method

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-

perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
.

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set

threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values


KFT

Command for volumetric Karl Fischer Titrations (

KFT

).

Principle

With this command volumetric titration for determining the water content by the
Karl Fischer method with automatic pre- and post-conditioning are carried out. A
typical feature of the Karl Fischer titration is the controlled reagent addition, which
depends on the control difference (current measured value – endpoint), with the

aim of reaching the measured value given as being the predefined target endpoint
as quickly and accurately as possible. Overtitrating is avoided to a great extent.
The titration termination at the endpoint takes place either drift-controlled or after a
waiting period. The volume added up to the endpoint gives the calculable reagent

consumption

Reagent addition takes place in three phases during the titration:

Initial addition

During this phase the dosing rate increases continuously. It starts with
the

Min. rate

and increases to the

Max. rate

.

Continuous addition

Addition is carried out at the

Max. rate

until the

Control range

is

reached.

Control range

In this range addition is finely controlled. Shortly before the endpoint is
reached addition is only carried out at the

Min. rate

.