Subwindow Rack data

755

swing head with beaker sensor must be installed and a suitable work posi-
tion with beaker contact must be defined for the lift, as this must move to

the beaker recognition position.
With the parameter

Beaker test

in the command

MOVE

you can determine

whether the determination is to be terminated if a beaker is missing with or
without the display of a corresponding message and whether the series is

to be continued or also terminated.

Rack offset

-5.00 ... [ 0.00 ] ... 5.00 °

The rack offset is a production-dependent tolerance value between the up-
per part of the rack and the lower part. The rack offset value is determined
during rack adjustment and shown here. It can be edited if necessary.

Lift positions

Tower 1

Defines the lift positions for Tower 1. These apply for all rack positions ex-
cept those that are defined as Special beaker.

Work position

[ 0 ] ... 235 mm

Work position for Lift 1. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.

Rinse position

[ 0 ] ... 235 mm

Rinse position for Lift 1. This lift position is used for rinsing the electrodes.

Shift position

[ 0 ] ... 235 mm

Shift position for Lift 1. Each time that the rack rotates the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here then the rotation will take place at
the current lift position. This means that the shift position must be selected

so that a safe rotation over the whole rack is possible at any time.