Subwindow Run
119
not yet been assigned in a command then the window
Device assign-
ment
appears for assigning the device.
• Check solutions
Checks whether the solutions demanded in the device-specific com-
mands are present and available.
• Check sensors
Checks whether the sensors demanded in the device-specific com-
mands are present and available.
Note
The Start test can also be triggered manually in the
READY
status with the menu
item
Tools, Run test
.
4 - Preconditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning
is switched on in the
START
command then pre-
conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (
Sample number
,
Statis-
tics
).
• Conditioning requirement not fulfilled
After the start of preconditioning the status changes to
COND BUSY
. In
this status conditioning can be started with
[Start]
even though the
conditioning requirement has not been fulfilled. Conditioning can also
be canceled with
[Stop]
or interrupted with
[Hold]
and then continued
with
[Cont]
. If an error occurs during conditioning then neither the Error
track nor the Exit track will be carried out.
• Conditioning requirement fulfilled
If the conditioning requirement has been fulfilled then the status
changes to
COND READY
. In this status the determination can be
started with
[Start]
. Conditioning can also be canceled with
[Stop]
or
interrupted with
[Hold]
and then continued with
[Cont]
. The condition-
ing requirement will continue to be monitored continuously until the
determination has been started.
5 - Start method
At the start of the method the following checks and actions are triggered:
• Check sample data
Checks that the entered Sample data corresponds to that defined in
the Method variables and is valid.
• Assign determination ID
An unambiguous determination ID is assigned for the determination.
• Reserve device (module)
The devices (or their functional units) used by the device-dependent
commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.
• Increment sample number
The
Sample number
is increased by
+1
.
• Increment statistics counter
If statistics has been switched on in both the
START
command and on
the tab
Single determinations
then the statistics counter is increased by
+1
. If a new method is loaded or if the statistics counter agrees with