Method
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(808, 809, 835, 836, 842)
[ 1 ] ... 3
(855)
Selection of the dosing device (exchange or dosing unit) to be used for
dosing the addition solution.
Solution
24 characters, solution name, [ not defined ]
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If
not defined
is selected no tests will be carried
out.
Note
If an existing solution is selected as the solution to be added then the
Concentra-
tion
and
Unit
of this solution will be adopted for calculating the standard addi-
tion. If
not defined
is selected then
Concentration
and
Unit
must be entered on
the
Standard addition
tab.
Sensor
Measuring input
[ 1 ] ... 2
(Titrando)
[ 1 ]
(855)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensorname, [ ISE electrode (Fluoride) ]
Selection of the ISE sensor from the list of sensors defined in the sensor ta-
ble.
Temperature measurement
Type of temperature measuring:
continuous
A temperature sensor must be connected. The temperature is measured
continuously.
[ automatic ]
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered
Temperature
(see
Meas-
uring parameters
) will be used.
off
The temperature will not be measured. The
Temperature
(see
Measuring pa-
rameters
) which has been entered manually will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.