Rs 232 – Metrohm tiamo 1.1 Manual User Manual

Page 726

background image

698

Configuration

Temperature correction

Shows the temperature correction set on the device. This parameter directly

affects the temperature control and allows it to be adjusted. In this way any
temperature difference that may occur between the oven and sample can
be compensated. The displayed temperature is the oven temperature cor-
rected by this value.

Gas

On the tab

Gas

the parameters for the gas flow to the 774 Oven Sample Proces-

sor are shown. These parameters cannot be edited; they are only read off by the
device.

Unit for gas flow

Shows the unit selected at the device for the gas flow display (

mL/min

or

L/h

).

Minimum gas flow

Shows the lower warning limit set on the device for the gas flow.

Maximum gas flow

Shows the upper warning limit set on the device for the gas flow.

Gas type

Shows the carrier gas.

Correction factor

Shows the correction for the gas flow measurement with "other" gases.

Note
If the minimum gas flow is undercut or the maximum gas flow is exceeded then
an error message will appear.

RS 232